ETT-KF787 卡爾費休水分測定儀(容量法)
特點:
中文顯示滴定過程,可進行中英文輸入、輸出。 彩色液晶屏顯示,圖形化界面,操作簡潔方便,易于掌握;觸摸屏輸入,只需輕輕點擊便能夠獨立完成繁雜的檢測過程。 全封閉設計的滴定臺,能自動更換溶劑,避免化學試劑與人體的接觸。 根據實驗的環境條件,可以設置“自動”或“手動”漂移值背景扣除,確保分析結果更為準確。 滴定結果可按GLP/GMP要求格式輸出,并對存儲的滴定結果進行統計分析,可連接中、英文微*型打印機,打印出結果報告。 隨機配有滴定監*控軟件,可監*控全部滴定過程,并通過該軟件進行版本升級。 可應用于制藥行業、食品、地質化工、石油化工、日用化工、農業、環保、水處理等諸多行業的質量檢驗或科研教學。
技術規格:
測量系統 測量范圍 極化電壓 0~±2500mv 極化電流 0~200uA 分辨率 極化電壓 0.01mv 極化電流 0.01uA
滴定管 1ml、5ml、10ml、20ml 分辨率 1/20000滴定管體積 絕對誤差(5ml滴定管以上≤3ul)
接口 2個USB口 2個RS232接口 2個TTL接口 1個Ethernet接口 1個CAN總線接口
性能 10ppm~100%含水量樣品 分辨率1ug水 精密度RSD<0.3%(5mg以上水)
=====================================================
Features:
Chinese display the titration process can be in both Chinese and English input and output. Color LCD display, graphical user interface, simple and convenient operation, easy to grasp; touch screen input, just a few clicks will be able to independently accomplish complex testing process. The fully enclosed design of the titration units, automatic replacement of the solvent, to avoid the chemical reagents and human contact. According to the environmental conditions of the experiment can be set to "automatic" or "manual" drift value of the background, net, and ensure more accurate results of the analysis. The titration results according to GLP / GMP requirements format output, and storage of titration results of statistical analysis, can be connected to the English mini-printer, print out a report on the results. Random with titration monitoring software can monitor all the titration process, and through the software upgrade. Quality testing or research and teaching can be applied to the pharmaceutical industry, food, geological, chemical, petrochemical, household chemicals, agriculture, environmental protection, water treatment and many other industries.
Technical Specifications:
Measurement system Measuring range Polarization voltage 0 ~ ± 2500mv Polarization current 0 ~ 200uA Resolution Polarization voltage 0.01mv Polarization current 0.01uA
Burette 1ml, 5ml, 10ml, 20ml Resolution 1/20000 of buret volume Absolute error (5ml burette above ≤ 3ul)
Interface 2 USB port 2 RS232 interface 2 TTL interface An Ethernet interface A CAN bus interface
Performance 10ppm ~ 100% moisture content samples Resolution 1ug water Precision RSD <0.3% (5mg above water)
============================================= |